ABSTRACT

The authors present a case in which macro-embolus from the ruptured atheromatous plaque developed during carotid artery stenting (CAS). A 63-year-old man who had suffered a left middle cerebral artery territory infarction had significant proximal left internal carotid artery stenosis required CAS procedure. Immediate after stent deployment, the patient showed abrupt neurological deterioration with 12 x 3 mm sized macro-embolus which was caught by the embolus protection device (EPD). Retrieval of the macro-embolus was performed safely and the patient recovered to pre-procedure status. Macro-embolus can be resulted during the CAS. The EPD can capture the macro-embolus and safe removal is technically feasible.


Subject(s)
Humans , Angioplasty , Carotid Arteries , Carotid Stenosis , Embolic Protection Devices , Embolism , Infarction , Middle Cerebral Artery , Stents